• The health care practitioner must monitor the patient's use of the
device to assure appropriate use and proper application of cold
therapy, including but not limited to application of an insulation
barrier between the patient's skin and the Cold Rush Pad.
• The product should never be left unattended when plugged in.
• Never drop or insert any object into any opening or hose.
• Do not operate where aerosol (spray) products are being used or
where oxygen is being administered.
• Do not attempt to drag or reposition the Cold Rush Cold Therapy unit
by pulling or tugging on the insulated hoses, as this may cause
damage to the unit and/or cause the hoses to disconnect. Use handle
when moving the Cold Rush Cold Therapy unit.
• If you experience disturbance in any other electrical equipment near
the Cold Rush unit, please try to unplug it from the wall socket to
verify that it is not the source of disturbance.

SAFETY STANDARDS

Cold Rush™ Cold Therapy System is tested and certified to comply with
the IEC60601-1 standard of electrical safety of medical devices and IEC/
EN60601-1-2, electromagnetic compatibility for medical electrical devices.
The company fulfills the requirements of ISO13485.

SETUP AND OPERATION

1. Ensure cold therapy unit is turned off, turn flow control knob
counterclockwise to the off position. Do not plug in to wall socket
(Figure 2)
2. Fill reservoir with water to the indicated fill line (Figure 3)
3. Add chopped or cubed ice to the reservoir up to the indicated fill line
(Figure 4)
4. Attach the provided 12V power supply to DC power port, do not plug
into wall circuit (Figure 5)
5. Apply insulation barrier over the patient's skin in the area to be
treated (Figure 6a)
6. Apply Cold Rush Pad over insulation barrier (not included). The Cold
Rush Pad should never be applied directly over the patient's skin
(Figure 6b)
7. Connect Cold Rush pad to hose extending from the unit. To ensure
proper connection, push couplings together until they snap into place
(Figure 7)
8. Plug the provided 12V power supply to wall socket (120V US, 240V
Europe) (Figure 8)
9. Turn flow control knob clockwise to the number "3" setting, this will
turn unit on (Figure 9)
10. Wait 10 minutes to allow temperature to stabilize before adjusting the
temperature for comfort (Figure 10)
11. Turn flow control knob counterclockwise to increase temperature.
12. Turn flow control knob clockwise to decrease temperature.
13. To turn the cold therapy unit off, turn flow control knob
counterclockwise until it stops.
14. Always turn the cold therapy unit off before disconnecting the Cold
Rush pad or any hoses. To disconnect the hoses, press the release
tabs on the coupling and pull the connectors apart.
15. Drain all of the water out of the cold therapy unit after each use.
16. Follow directions 2–10 before each use.