verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or
relocating the Pulse Oximeter.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should
be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the Pulse
Oximeter
The Pulse Oximeter is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the Pulse Oximeter as recommended below,
according to the maximum output power of the communications
equipment.
Rated
maximum
output
power of
transmitter
(W)
0.01
0.1
1
10
100
Separation distance according to frequency of
transmitter
(m)
150KHz to
80MHz
3
5 .
d
P
E
1
0.12
0.37
1.17
3.69
11.67
80MHz to
800MHz
3
5 .
d
P
E
1
0.12
0.37
1.17
3.69
11.67
800MHz to
2.5GHz
7
d
P
E
1
0.23
0.74
2.33
7.38
23.33