2
2
2
5
6
Connecting the AC Adapter
(Optional); not for YCO14-Y
when Installed in Zone 2 or
Division 2 Potentially Explosive
Areas
2
t
t
t
8
EVO1X | LAB1X Installation Instructions
Fasten the
t
strain relief
YSR01 to the
converter (1)
using the two
screws (2).
1
Attach the
t
USB cable to
the strain
2
relief (1)
using the
3
clamp (3)
and the two
1
screws (4).
Use the
t
protective
cap (6) to
seal up the
left-hand
USB port (5).
Insert an additional USB cable
(2) into the ex-link converter.
Insert the USB cable into the
YEPS01-USB AC adapter.
Plug the AC adapter into a wall
outlet (supply voltage).
3.7
Anti-theft Locking Device
t
3.8
Leveling
Leveling for PMA.HD and Verified Models
Leveling the scale compensates for slant or unevenness at the
place of installation. The scale must be perfectly horizontal to
ensure consistent, reproducible weighing results.
The scale needs to be re-leveled and then adjusted if
necessary each time its setup location is changed.
t
12*
9*
− Air bubble at "12 o'clock:" Turn
6*
− Air bubble at "3 o'clock:" Turn
* o'clock
− Air bubble at "6 o'clock:" Turn
− Air bubble at "9 o'clock:" Turn
3.9
Warm-up Time
Using a Scale Verified for Use in Legal
Metrology:
Ensure that there is a warm-up time of at least
24 hours after initial connection to the power
supply.
To ensure accurate results are delivered,
0
the scale must warm up for at least
30 minutes after initial connection to
the power supply.
Only then will the device have reached the
required operating temperature.
30
If required, secure the scale at
the back.
Turn the two leveling feet as
shown in the illustration until
the air bubble is centered within
the circle of the level indicator.
both feet clockwise.
the left foot clockwise and the
right foot counterclockwise.
both feet counterclockwise.
the left foot counterclockwise
and the right foot clockwise.