ACCULAN drill GA 620 D
5.3.1 Coupling
Push the tool chuck 3, 6, 8, 9, 10 or 11 onto the driven shaft of
the drill until it is clearly fixed into place (a snap action).
Fig. 8: Coupling
Regarding the chuck 6, 8, 9, 10 and 11, it may be necessary to turn
the tool adapter slightly on the shaft during coupling to enable the
octagonal section of the driven shaft to engage.
When coupling, do not hold the release sleeve 4.
To attach the Kirschner wire chuck 12 or 14, fit the Kirschner wire
chuck on the drill and secure it with the clamp lever 13.
5.3.2 Uncoupling
The chuck 3, 6, 8, 9, 10 or 11 are removed by first pressing them
gently against the drill and then removing them by pulling back the
release sleeve 4.
Fig. 9: Uncoupling
To remove the Kirschner wire chuck 12 or 14, open the clamp
lever 13 and pull the Kirschner wire chuck off the drill.