Guidance and manufacturer's declaration - electromagnetic emissions
The electrical breast pump NUK e-MOTION is intended for use in the electromagnetic environment specified
bellow. The costumer or the user of the electrical breast pump NUK e-MOTION should assure that it is used
in such an environment.
Immunity test
IEC 60601-1 test level
Electrostatic
+/- 6kV contact
discharge (ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power supply
transient/ burst
lines
IEC 61000-4-4
+/- 1 kV for input/ output
lines
Surge
+/- 1 kV differential mode
IEC 61000-4-5
+/- 2 kV common mode
Voltage dips, short
<5% for 5s
interruptions and
<5% for 0,5 cycles
voltage variations on
40% for 5 cycles
power supply input
70% for 25 cycles
lines
IEC 61000-4-11
Power frequency
3 A/m
(50/60Hz) magnetic
fields
IEC 61000-4-8
Compliance level
Electromagnetic environ-
+/- 6kV contact
Floors should be wood, con-
crete or ceramic tile. If floors
+/- 8 kV air
are covered with synthetic
materials, the relative humidity
should be at least 30%.
+/- 2 kV for power supply
Mains power quality should be
lines
that of a typical commercial or
hospital environment.
+/- 1 kV for input/ output
lines
+/- 1 kV differential mode
Mains power quality should be
that of a typical commercial or
+/- 2 kV common mode
hospital environment.
<5% for 5s
Mains power quality should be
that of a typical commercial or
<5% for 0,5 cycles
hospital environment. If the
40% for 5 cycles
user of the electrical breast
70% for 25 cycles
pump NUK e-MOTION
requires continued operation
during power mains interrup-
tion, it is recommended that
the electrical breast pump NUK
e-MOTION be powered from
an uninterruptible power sup-
ply or battery.
3 A/m
Power frequency magnetic
fields should be at levels charac-
teristic of the typical location in
a typical commercial or hospital
environment.
117
ment - guidance