4.2 Frequency select
Set the
the specimen to be test. Generally, the electrolytic capacitor is set to
120 Hz. Others are set to 1 kHz in general.
4.3 Parallel / Series mode
Set the
circuit mode. However, in the specimen having a high impedance in
general, measurement is made in "parallel equivalent circuit" mode
"PAL",
while
measurement is made in "series equivalent circuit" mode "SER".
Note:
For resistances (R) is only available the parallel (PAL) mode.
4.4 RANGE Button
Press
turn off the "AUTO" enunciator. (The meter remains in the range it
was in when manual ranging was selected).
In the Manual Range mode. Each time press
the range (and the input range enunciator) increments, and a new
value displayed. To exit the Manual Range mode and return to
autoranging, press and hold down
The "AUTO" enunciator turns back on.
Page 18
[6] button switch to 120 Hz or 1 kHz according to
[7] button switch to parallel or series measuring
in
the
specimen
[4] button to select the Manual Range mode and
INSTRUCTION MANUAL. MZ-505C
having
a
[4] button for 2 seconds.
low
impedance,
[4] button,
04/2015