Range Limit
The range limit is the distance at which measurements are ignored. Generally this
refers to the bottom of the container being measured. If a measurement is
detected beyond the range limit it results in a Loss Of Echo (LOE) reading. The
result of this reading is determined by the Fail-Safe Mode, see page 17 for more
information.
Factory setting: 5.50 m (18.0 ft)
range
limit
0 to 3 m vessels (solids)
0 to 5 m (liquids)
Speed of Response
The speed of response adjustment allows the user to collectively set a number of 
operating parameters.
measurement
response:
agitator
discrimination:
filter:
fail-safe timer:
SP
measurement response
1
0.3 m / min (0.1 ft / min)
2*
1 m / min (3.3. ft / min)
3
5 m / min (16.4 ft / min)
4
* Factory setting
Page EN-16
is the limit to which the Pointek ULS200 is able to keep up with
rates of change.
If the Pointek ULS200 measurement cannot keep up with the
rate of level change, set the adjustment from 1 to 2. 
If the Pointek ULS200 still cannot keep up with the rate of 
level change, set the adjustment option to 3. Avoid 
choosing an option that is too fast for your application.
discriminates between agitator blades in motion 
and the material (target) surface.
discriminates between false echoes from acoustical 
and electrical noise and the material (target) surface.
establishes the `Waiting' period from the time a loss
of echo or operating fault condition starts until the
fail-safe default is effected.
immediate
Pointek ULS200 – QUICK START MANUAL
range
limit
greater than 3 m vessels (solids)
greater than 5 m (liquids)
agitator
filter FLS timer
discrimination
on
on
on
on
on
on
off
off
10 min
10 min
3 min
3 min