6 CALIBRATION (continued)
To manually enter the calibration tile data:
1
Press Menu/Setup/Calibration/Tile Setup/Manual Setup
é
2
Use the
é
3
Use the
to save or 'Escape' to exit and disregard any amendments made.
4
Repeat steps 2-3 to enter the calibration data for 60° and 85°
angles.
é
5
Use the
press 'Select'.
Use the çè softkeys to enter the tile serial number.
6
7
Select 'Ok' to save the changes or 'Escape' to exit and disregard
any amendments made.
The calibration tile information can be viewed at any time via
Menu/About/Tile Information.

6.2 CALIBRATING THE GAUGE

The Elcometer 480 will calibrate automatically on switch on (if enabled)
and/or calibration can be performed at any time.
To enable auto calibration (T):
1
Press Menu/Setup/Calibration.
é
2
Use the
'Select'.
The calibration procedure will now commence automatically each time
the gauge is switched on and a calibration tile is attached to the base
of the gauge.
To manually calibrate the gauge:
1
Attach the Elcometer 480 calibration tile to the base of the gauge
or place the gauge on to an independent gloss tile.
2
Press Menu/Calibrate (B & T) or Batch/New Batch/Calibrate
when in Batching (T).
Calibration details for each individual batch can be viewed at any time
via Menu/Batch/Review Batch/Calibration Information (T).
Model dependent: Dual and Triple Angle models only.
en-13
R
softkeys to highlight 'Set 20° ' and press 'Select'.
ê
softkeys to enter the required value and press 'Set'
ê
softkeys to highlight 'Set Tile Serial Number' and
ê
softkeys to highlight 'Auto Calibrate' and press
ê
www.elcometer.com