Integra
Luxtec
®
DLX UltraLite
Pro Camera System
®
White Balance
White balance optimizes the color definition at different illumination levels.
White Balance before each procedure is strongly recommended.
1. Fully open the iris of the headlight module.
2. Point the camera and headlight at a white object.
Make sure the image fills the monitor entirely.
3. Press the "AWB" button on the rear of the CCU and hold for 5 seconds.
The camera will white balance automatically.
Troubleshooting
Problem
No image on monitor
Poor Image quality
Image not centered on monitor
Colors not correct
Black and white image
Dim image
If you are still having issues with your system after following the above troubleshooting tips, please contact your local Integra sales representative
8
®
Cause
Camera not turned on
Camera head not connected to controller
CCU not plugged into AC outlet
Monitor not turned on
Monitor settings not correct
Intermediate component (recording device/
printer) connection not correct
Components not connected properly
Dirty camera lens / obstruction
Bad / damaged camera
Video standard incompatibility (NTSC/PAL)
Illumination spot not vertically centered on
video image
White balance is not correct
S-Video cable connection is loose
Insufficient light at surgical site
Action
Turn power on
Connect camera head cable to front of CCU
with power OFF
Plug CCU power cord into outlet
Turn monitor on
Consult monitor manufacturer's guide
Consult connection diagram in this manual
Consult connection diagram in this manual
Refer to cleaning instructions
Contact your Integra distributor
Consult components' manufacturers' guides
Contact your Integra distributor
Follow white balance instructions in this
manual to reset
Secure S-Video cable connections
Increase light output per light source user
instructions