Integra
Luxtec
®
DLX UltraLite
Pro Camera System
®
Electromagnetic Immunity
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
Immunity against
electrostatic discharge,
ESD (IEC 61000-4-2)
electrical fast
transient/bursts
(IEC 61000-4-4)
surges on AC mains lines
(IEC 61000-4-5)
power frequency
magnetic field 50/60 Hz
(IEC 61000-4-8)
voltage dips and short
interruptions on AC
mains input lines
(IEC 61000-4-11)
12
®
IEC 60601-1-2 test level
contact discharge:± 6 kv
air discharge: ± 8 kv
power supply lines: ± 2 kv
longer input/output lines: ± 1 kv
common mode: ± 2 kv
differential mode: ± 1 kv
3 A/m
dip >95%, 0.5 periods
dip 60%, 5 periods
dip 30%, 25 periods
dip >95%, 5 seconds
Compliance level
Electromagnetic environment
(of this device)
± 6 kv
Floors should be wood, concrete or ceramic tile.
± 8 kv
If floors are covered with synthetic material,
the relative humidity should be kept at levels to
reduce electrostatic charge to suitable levels.
± 2 kv
Mains power quality should be that of a typical
± 1 kv
commercial or hospital environment.
± 2 kv
Mains power quality should be that of a typical
± 1 kv
commercial or hospital environment.
3 A/m
Equipment which emits high levels of power
line magnetic fields (in excess of 3A/m) should
be kept at a distance to reduce the likelihood of
interference
>95%, 0.5 per.
Mains power quality should be that of a typical
60%, 5 per.
commercial or hospital environment. If user
30%, 25 per.
requires continued operation during power
>95%, 5 sec.
mains interruptions, insure that batteries are
installed and charged. Insure that battery life
exceeds longest anticipated power outages
or provide an additional uninterrupted power
source.