This equipment is intended for use in the electromagnetic environment specified below.
The customer or user of this equipment should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
150 kHz to 80 MHz
rf coupled into lines
outside ISM bandsa
(IEC 61000-4-6)
radiated rf
150 kHz to 80 MHz
(IEC 61000-4-3)
in ISM bands
80 MHz - 2.5 GHz
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
equipment.
Compliance level
3 Vrms
10 Vrms
a
10 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment,
including cables, should be used no closer to any
part of the DLX UltraLite® Pro Camera Controller
Unit than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter as below:
Recommended separation distance
d=1.2/V1]√P
d=1.2/√P
80 MHz to 800 MHz
d=2.3/√P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol: