Problem
Pump has no suction
• Lid 4 is not fitted correctly
or too little suction
• Overflow protection 5 is wrongly
inserted
• Suction regulator dial 1 is not fitted
correctly
• Pump unit is not fitted correctly:
intake fitting on lid 4 is not fitted
correctly in the suction connector
on the pump unit 2
• Lip valve is incorrectly positioned
• Pump is being held at an angle
Pump lever feels stiff • Too much friction on the piston
seal
• Overflow protection 5 is wrongly
oriented
Suction regulator
• Too much friction on the suction
feels stiff
regulator
Medical product in accordance with 93/42/EEC
Temperature range
– Operation
– Storage
13. Troubleshooting
Reason
• Ensure that the lid has been fully pressed on
(see Section 8.).
• Check that the overflow protection is correctly
positioned and aligned.
• Pull the dial off, rotate slightly, and press back
on.
• Remove pump unit and replace. Check that the
intake fitting on the lid 4 is fully inserted into
the suction connector on the pump unit 2 (see
Section 8.).
• Position the lip valve correctly
(see Section 8.).
• Hold the pump as upright as possible.
• Remove pump unit 2. Press hand lever as far as
possible. Apply a little baby oil to the piston
seal.
• The connecting piece must point towards the
suction funnel (see Section 8.).
• Pull the dial off. Apply a drop of baby oil to the
suction regulator. Press the dial back on.

14. Technical Data

Class I
6 °C to +40 °C
0 °C to +40 °C