6. Alarms and troubleshooting
A description of the device alarms and indicators is provided.
Use this section as a reference for your device alarms. If the alarm does not
stop, contact your healthcare provider to inform them of the alarm condition.
If an alarm occurs:
1. The status indicator illuminates yellow.
2. An alarm screen will display.
3. The audible alarm sounds every 20 seconds.
Some alarms allow the audible alarm to be paused for approximately
2 minutes. The Low Battery alarm allows the audible alarm to be paused for
approximately 15 minutes. If the cause of the alarm is not resolved within this
time the alarm will sound again. If the audible alarm has been paused and a
new alarm state occurs, the audible alarm will sound, and the touchscreen
will display the new alarm. When multiple alarm states are present, the
device will alternate between alarm screens every 5 seconds.