Troubleshooting Guide; Over Vacuum Alarm; High Vacuum Alarm - smith&nephew 66801280 Manual Del Usuario

Ocultar thumbs Ver también para 66801280:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 369
EN

Troubleshooting guide

Over Vacuum Alarm

The device vacuum exceeds a pressure of 235mmHg and the device has
stopped therapy.
Troubleshooting
1. Power Off and restart device.
2. If alarm recurs there is a potential malfunction of device.
Contact your healthcare provider.

High Vacuum Alarm

The device vacuum level is higher than the prescribed therapy and the
device has stopped therapy.
Troubleshooting
1. Power Off and restart device.
2. If alarm recurs there is a potential malfunction of device.
Contact your healthcare provider.
• Status indicator illuminates yellow.
• Audible alarm sounds every
20 seconds.
• Audible alarm cannot be paused.
• Status indicator illuminates yellow.
• Audible alarm sounds every
20 seconds.
• Audible alarm cannot be paused.
28

Publicidad

Tabla de contenido
loading

Tabla de contenido