PRODUCT DESCRIPTION
The dual channel Dupel Iontophoresis System delivers ionic drug solutions directly to
the body site being treated and offers:
Two channels for treatment of two different sites simultaneously.
•
Automatic calculation of time based on set dose and amplitude.
•
Self diagnosing capability that alerts the user when the circuit is not complete.
•
System components and accessories may include:
Dupel Iontophoresis Device
•
User's Manual
•
Empi Buffered Iontophoresis Electrodes
•
Dupel Iontophoresis Lead Wires
•
Batteries
•
Controls And Features
1. Channel Output Jacks connect lead wires to the device. Channels 1 and 2 are
independently controlled by Current Knobs 1 and 2.
2. ON/OFF Indicator Lights illuminate continuously
when the corresponding channel is delivering
current. They flash when there is an interruption of
current flow on the corresponding channel.
3. Current Knobs control the current output of Channel
1 and 2 between 0.5 and 4mA.
4. Treatment Mode Switch has four settings.
CH1 Set Up Mode, (S1), allows the target dose for
•
CH1 to be set.
CH2 Set Up Mode, (S2), allows the target dose for
•
CH2 to be set.
Pause Mode, (P), pauses the output during a
•
treatment.
Run Mode, (R), initiates or resumes treatment.
•
5. Power Switch turns the device ON or OFF.
6. Display Button selects which parameter is
shown in the display window, either Time (T),
Dose (D) or Current (C).
7. Dosage Buttons (∆ ,∇) set the dosage
(mA
min) for each channel. The Treatment Mode Switch
•
must be set to (S1) or (S2) to set dosage for CH1 or CH2 respectively.