Stress System Setup
To setup the Tango+ with your stress system, please visit the Tango+ Interface Notes
on the SunTech Medical website at www.SunTechMed.com. Select Customer Service
► Downloads and under the Products section, select Tango+. If your stress system is
not listed, please contact:
To test the stress system setup with Tango+, you will need to hook up a patient with
ECG leads and take a blood pressure measurement. You cannot use a heart rate or
blood pressure simulator to test whether the Tango+ is working with your stress system.
The Tango+ monitor requires that the ECG signal and the Korotkoff sounds, collected
by the microphone in the cuff, originate from the same source, meaning the patient.
DISCLAIMER: If the stress ECG system manufacturer modifies the
communication protocol with the Tango+ monitor without informing
SunTech Medical, you may experience difficulties communicating with
the stress system.