IRP25.50
2.3
Battery Replacement
Risk of damage to the equipment!
•
Before opening probe, clean and dry well!
•
Do NOT blow off with compressed air!
•
Replace empty battery immediately!
1.
Laterally pull the snap ring off from the probe (1./Fig. 12).
2.
Carefully slide the battery compartment sleeve down (2./Fig. 12).
The touch probe IRP25.50 ist protected against wrong polarity. Therefore there is no need to pay attention to the
polarity when inserting a new battery!
3.
Take the used battery off the battery compartment and insert new battery (3./Fig. 12).
Risk of damage to the equipment!
•
When closing battery compartment, ensure the O-Ring is properly seated!
4.
Carefully slide the battery compartment sleeve upwards (4./Fig. 12).
5.
Position the snap ring at the probe laterally and slide it over the probe (5./Fig. 12).
After changing the battery, the time for starting the touch probe (reboot) is approx. 5 s.
1.
Snap ring
(9 V, 1200 mAh, lithium)
(5525)
Fig. 12
Battery replacement
NOTICE
INFORMATION
NOTICE
INFORMATION
2.
Battery
(5677)
3.
O-Ring 53,5x1,2 Viton
(1477)
Battery compartment
sleeve
Operation
4.
5.