Battery Replacement - Hexagon m&h IRP25.50 Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

IRP25.50
2.3

Battery Replacement

Risk of damage to the equipment!
Before opening probe, clean and dry well!
Do NOT blow off with compressed air!
Replace empty battery immediately!
1.
Laterally pull the snap ring off from the probe (1./Fig. 12).
2.
Carefully slide the battery compartment sleeve down (2./Fig. 12).
The touch probe IRP25.50 ist protected against wrong polarity. Therefore there is no need to pay attention to the
polarity when inserting a new battery!
3.
Take the used battery off the battery compartment and insert new battery (3./Fig. 12).
Risk of damage to the equipment!
When closing battery compartment, ensure the O-Ring is properly seated!
4.
Carefully slide the battery compartment sleeve upwards (4./Fig. 12).
5.
Position the snap ring at the probe laterally and slide it over the probe (5./Fig. 12).
After changing the battery, the time for starting the touch probe (reboot) is approx. 5 s.
1.
Snap ring
(9 V, 1200 mAh, lithium)
(5525)
Fig. 12
Battery replacement
NOTICE
INFORMATION
NOTICE
INFORMATION
2.
Battery
(5677)
3.
O-Ring 53,5x1,2 Viton
(1477)
Battery compartment
sleeve
Operation
4.
5.
73

Publicidad

Tabla de contenido
loading

Productos relacionados para Hexagon m&h IRP25.50

Tabla de contenido