IRP25.50
2.8
Adjusting Trigger Force (Measuring Unit PP41.00 only)
If there are strong vibrations in the machine or if heavy probe inserts are used, this can lead to an increase in the
probing force to improve the measurement results.
1.
Unscrew the stylus (refer to chapter 2.2.3).
2.
Increase or decrease the probing force with a Hexagon key AF1.3 mm (refer to Fig. 18).
Hexagon key AF1.3 mm
Fig. 18
Adjusting trigger force
3.
Screw in the stylus (refer to chapter 2.2.3) and align the stylus to spindle center (refer to chapter 2.4).
4.
Calibrate the probe.
INFORMATION
Factory setting
(0227)
Revolutions based on lowest
force setting
min. 0
1
2
3
4
5
6
7
8
9
max. 10
+
-
Operation
≈ Probing force with 50 mm stylus in N
Z
X/Y
2,5
0,3
3,5
0,4
4,5
0,5
5,5
0,6
6,5
0,7
7,5
0,9
8,5
1
9,5
1,1
10,5
1,2
11,5
1,3
12,5
1,4
79