Measuring Unit Replacement

1.
Remove the stylus or the temperature sensor respectively according to chapter 2.2.3 and 2.2.1.
2.
Remove the battery from the touch probe according to chapter 2.3.
3.
Insert the C-Spanner into the bores of the measuring unit and carefully unscrew the measuring unit in counter
clockwise direction.
Risk of damage to the equipment!
When inserting the new measuring unit, ensure that the O-Rings inside the touch probe housing and on the
measuring unit are properly seated!
4.
Insert the new measuring unit into the touch probe and carefully screw it in by hand.
5.
Insert the C-Spanner into the bores of the new measuring unit screw the measuring unit in completely to the
stop.
6.
Mount the stylus or the temperature sensor respectively according to chapter 2.2.3 and 2.2.1.
7.
Reinsert the battery into the touch probe according to chapter 2.3.
After inserting the battery, the time for starting the touch probe (reboot) is approx. 5 s.
8.
When using PP41.00 or MY21.00 align the stylus to the spindle center.
9.
Calibrate the probe.
Measuring unit
PP41.00
Fig. 15
Measuring unit replacement (Example: PP41.00 -> TP44.10)