Descargar Imprimir esta página

Danger Of Burns - Erbe 20132-106 Manual Del Usuario

Apc-aplicadores

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 72
OBJ_DOKU-189911-006.fm Seite 26 Donnerstag, 17. Juni 2021 9:31 09
This is ensured if:
• the APC applicator with several black rings is visible through
the endoscope at the distal end of the second ring.
• the APC applicator with just one black ring is visible through
the endoscope at the distal end of this ring.
3.9

Danger of burns

Argon plasma can strongly heat electrically conductive materials,
e.g. metal. To avoid burns, never direct argon plasma towards
electrically conductive materials.
3.10 Risk of accidental tissue damage with the APC 300
If the gas bottles of the APC 300 are not opened or are com-
pletely empty, a spray coagulation is performed with the instru-
ment after each activation. There is an audible warning signal;
the bottles on the display of the APC 300 blink and an error mes-
sage appears. In the spray coagulation mode there is a danger
of unintentional tissue damage; the instrument can also be dam-
aged!
Before using the APC 300 always open the gas bottles. Observe
the self-check performed by the unit. How much gas is there left
in the bottles? The bottle symbols on the display of the APC 300
give an approximate indication of the quantity of gas remaining.
Never work with two gas bottles which are almost completely
empty. When one gas bottle is empty, the APC 300 switches over
to the second gas bottle automatically.
Always observe the warning signals and error messages.
26

Publicidad

loading

Este manual también es adecuado para:

20132-110