If damaged, do not use this product!
APC-applicators 20132-106, 20132-107, 20132-109, 20132-
110, 20132-111, 20132-112, 20132-113, 20132-114, 20132-
115, 20132-116: These APC-applicators are attached to con-
necting cable 20132-046.
For APC-applicator 20132-106 only: In close anatomical rela-
tionships, the protection tongue 20132-129 reduces the risk of
unintentional coagulation of tissue that lies opposite to the ap-
plication direction.
Check whether the plasma ignites upon first activation (see
chapter "Ignition test").
Protect the APC-applicator against physical damage. Do not
bend! Do not throw! Do not use force!
6
Ignition test
If the APC-applicators are stored for a longer period of time, it is
possible that the first ignition of the plasma is more or less de-
layed. For this reason, carry out an ignition test before the first
application of the Argon Plasma Coagulation. To do this, use the
ignition adapter (Art.-No. 20183-072). To perform an ignition
test follow the instructions in the Notes on Use for the ignition
adapter.
7
Cleaning, Disinfection, Sterilization
7.1
Safety instructions for processing
Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
Suitable for a maximum of 20 ultrasonic cycles.