4

Recommended Unit Settings

4.1
For APC 300 in combination with a ICC 200, 300 or
350
• ICC 200: Argon Mode: 10 – 40 W
• ICC 300: Spray Coag: 10 – 40 W
• ICC 350: Spray Coag: 10 – 40 W
• APC 300: Argon flow is preset by instrument recognition.
4.2
For APC 2 in combination with a VIO
electrosurgical unit
After being connected to the APC 2, the instrument is detected
automatically and the instrument-specific display with basic set-
tings for the operating modes CUT and COAG is displayed on the
VIO electrosurgical unit.
If individual settings have been saved to the program for the op-
erating parameters mode, effect, rated power and/or argon
flow, these are displayed. Use the display on the VIO electrosur-
gical unit to make sure that the instrument has been correctly
detected.
5

How to Use

Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
Before each use, check the insulation, especially the ceramic in-
sulation at the distal end of the APC-applicator, for damage.
43