1. Roll up the product loosely and lay it in a suitable rinsing
basket. In doing so make sure the product is not touching
other instruments or parts of instruments.
2. Connect the product via the irrigation adapter to a rinsing
station.
3. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection: 5 to 10 minutes at 90 to 95 °C,
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
4. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
5. Remove the irrigation adapter when the program has ended.
9.8

Check

1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering, discoloration.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
– Bending of the product or parts of the product. Do not at-
tempt to straighten bent instruments!
If damaged, do not use this product!
9.9

Packaging

1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.
43