The instrument satisfi es the requirements for electromagnetic
compatibility. Please note that under the infl uence of unfavourab-
le fi eld strengths, e.g. during the operation of wireless telephones
or radiological instruments, adverse effects on function cannot
be excluded.
Attention!
There is a possible danger of infl ammation of gases, if the inst-
rument isoperated in the presence of infl ammatory mixtures or
mixtures of pharmaceuticals and air or oxygen or laughing gas!
Never attempt to take the instrument apart! There is a danger of
life-threatening electrical shock. Unplug the instrument before
cleaning or when disinfecting.
For devices of protection class I : „Attention : To avoid the risk of
an electric shock the device must only be connected to the public
power supply with a protective earthing conductor".

3. Intended use

The Wall /Mobile, Anesthesia device ri-former
these instructions for use was manufactured for use with various
instrument heads and modular components for non-invasive
diagnostics.

4. Attachment

a.) Drilling instructions/drilling plan
The drilling instructions and the drilling plan are enclosed sepa-
rately. Follow the drilling instructions in order to drill the holes in
the wall.
b.) Attaching the wall mounting plates
After you have drilled the holes, take the plugs supplied and push
them into the holes as far they will go.Take the wall mounting
plate and hold it onto the wall so that the screws can be pushed
through the holes of the mounting plate into the plugs. Now
screw in the screws with a screw driver, as far as they will go.
c.) Attachment of the diagnostic station
When all screws have been screwed in tightly, take the diagnostic
station and guide the screw heads through the openings (1). Then
press the diagnostic station downwards until it snaps into place.
d.) Attachment of the extension module
Connect the diagnostic station and the extension module with the
help of the connecting cable. In order to plug in the connecting
cable, remove the sliding cover (2) ofthe diagnostic station.
Close the casing opening of the extension module, which is not
needed, with the sliding cover (2). Take the extension module and
guide the screw heads through the openings (1). Then press the
extension module downwards.
Follow the instructions in the operation
manual
Rel. Humidity
Used electrical and electronic devices
should not be disposed of with normal
household waste, but should be disposed of
separately according to national or
EU guidelines.
®