This ME device is intended exclusively for use by professional
medical staff. This device can cause radio interference and can
disrupt the operation of equipment nearby. Suitable remedial
measures, such as for instance re-alignment, re-arrangement of
the ME device or shielding, can become necessary.
Guidelines and manufacturer's declaration - electromagnetic emissions
The Diagnostic station ri-former
environment as specified below. The customer or the user of the Diagnostic station
ri-former
Emission measu-
rements
HF emissions
according to
CISPR 11
HF emissions
according to
CISPR 11
Harmonics
emissions
according to
EC61000-3-2
Voltage fluctuation
/ flicker
emissions accor-
ding to
IEC61000-3-3
18
®
is intended for operation in an electromagnetic
®
should ensure that it is used in such an environment.
Compliance
Group 1
Class B
Not applicable
Not applicable
Electromagnetic environment -
guidelines
The Diagnostic station ri-former
employs HF energy solely for an
internal function. Its HF emission is
therefore very low and it is unlikely
that neighboring electronic devices
will be affected by interference.
The Diagnostic station ri-former
is intended for use in all facilities,
including living quarters and such
as are directly connected to a public
power supply that also supplies
buildings that are used for
residential purposes.
®
®