Welch Allyn CP Serie Manual Del Usuario página 26

Ocultar thumbs Ver también para CP Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 91
58
Chapter 5 Performing ECG Tests
Figure 38. Auto ECG Acquisition" Screen
Doe, Jane
Auto ECG Acquisition
I
V
aVR
Leads
4. Verify ECG quality on the screen.
5. (Optional) Adjust the waveforms.
See
6. Press Print ECG.
7 .
If prompted, choose whether to wait for the electrocardiograph to acquire 10 seconds
of filtered, processed data before printing.
If you override the wait time and print the available data immediately, be aware that
the printed data will be insufficient in quality or quantity or both.
The report prints.
Note
After printing, the Auto ECG Post-Print screen appears.
9:17AM Oct 16 05
AUTO
ON
ON
Baseline
Muscle
Gain
Filter
Filter
WARNING Do not perform ST segment analysis on the ECG screen display,
since these ECG representations are scaled. Make manual measurements of
ECG intervals and magnitudes on printed ECG reports only.
"Adjusting the ECG Waveforms"
If you do not enter the patient data, write the patient identification
information on the printed report immediately after the ECG test to avoid
associating the ECG report with the wrong patient.
If a red stripe appears along the edge of your report, replace the paper.
See
"Loading the Thermal Chart Paper"
Welch Allyn CP 200 Electrocardiograph
HR 120
Softkeys for adjusting the
waveforms or printing
Print
ECG
on page 68.
on page 23.
Figure 39
on page 59.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cp 100Cp 200

Tabla de contenido