Disconnecting The Product From The Unit; Precleaning The Product In The Procedure Room - Erbe 20416-032 Manual Del Usuario

Criosonda flexible
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 46
OBJ_DOKU-181800-011.fm Seite 24 Dienstag, 16. März 2021 8:28 08
2. Place the cryotip on the target tissue or foreign object apply-
ing light pressure and activate the cryosurgical unit's freeze
function.
3. While keeping the cryosurgical unit's freeze function perma-
nently activated, perform the following steps:
– Remove the tissue or the foreign body by withdrawing
the instrument together with the endoscope from the
operating field. Do not rotate the cryo instrument.
– Withdraw the endoscope together with the instrument.
4. To collect the sample, deactivate the cryosurgical unit's
freeze function and allow the instrument to thaw completely
(e.g. dipped in a sample container filled with liquid).
5. Remove the instrument from the endoscope. Make sure that
the angle of the endoscope is not too steep and the probe is
not clamped.
6.5

Disconnecting the product from the unit

1. Unscrew the cryoprobe from the connecting adapter (Art. No.
20416-034).
2. Screw the protective cap onto the probe coupling.
6.6

Precleaning the product in the procedure room

1. Use a soft cloth (such as a compress) to remove any obvious
contamination immediately after use and/or place the prod-
uct in deionized water.