7
Cleaning, disinfection, sterilization
7.1
Safety instructions for processing
May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
The probe coupling must remain sealed during the entire prepa-
ration process with the protective cap. Moisture penetrating into
the tubing system causes malfunctions.
Only roll up the application element loosely; do not kink or crimp
it.
Only roll up the connection hose loosely; do not kink or crimp it.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
7.2
Processing limitation
The flexible cryoprobe 20416-032 must be tested in each case
after 100 uses/sterilizations, but at least once per year.
The flexible cryoprobe 20416-037 must be tested in each case
after 30 uses/sterilizations, but at least once per year.
Contact Erbe Elektromedizin or your local sales partner for the
test.