WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the instrument.
BiClamp with MF connector (20195-xxx):
1. Connect the instrument to the MF socket on the VIO 300 D or
the VIO 3.
2. Check whether the instrument has been detected: On the
VIO 300 D, "BiClamp" mode is displayed and on the VIO 3,
the "BiClamp" instrument is shown.
If a correctly connected instrument is not detected, replace
the instrument.
If the error persists, replace the electrosurgical unit.
Note: You can also connect the instrument to an MF-2 socket on
a VIO 300 D or an MF-U socket on a VIO 3 using adapter 20183-
101.
BiClamp with MF-2 connector (21195-xxx):
1. On the VIO 300 D: Connect the instrument to an MF-2 socket.
On the VIO 3: Connect the instrument to an MF-U socket.
MF
MF-2
MF-U