6. Technical data and operating conditions
6.1

Technical data

Power supply
Overvoltage category
Safety class
Contamination level
Acceptable voltage fluctuations
Max. power consumption
Max. current consumption
(informative)
Acceptable data for pumps of
other manufacturers:
Max. output, final vacuum
Values and dimensions of electrical
fuses
Dimensions of the closed furnace
Usable size of the firing chamber
Max. firing temperature
Weight
10
110 – 240 V~ / 50/60 Hz
II
I (The device requires a ground connection
via the mains connection)
2
± 10%
1650 W
16 A at 110–130 V~
9 A at 200–240 V~
250 W /max. leakage current 0.75 mA
< 50 mbar
Heating circuit:
T15AH 250 V
6.3 x 32 mm
Vacuum pump:
T5AH 250 V
5 x 20 mm
Depth: 494 mm
Width: 280 mm / 415 mm (without/with
cooling tray)
Height: 450 mm / 573 mm (closed/open)
Diameter: 55 mm
Height: 40 mm
1 560°C (2192°F)
28.5 kg

Applied standards

The furnace was tested in accordance with the following standards:
– EN 61010-1:2010/A1:2019
Device safety
– UL 61010-1:2012/R:2018-11
– CAN/CSA-C22.2 No. 61010-1:2012/A1:2018-11
– IEC 61010-1:2010+AMD1:2016
– EN 61010-2-010:2020
– UL 61010-2-010:2019
– CAN/CSA-C22.2 No. 61010-2-010:2019
– IEC 61010-2-010:2019
– EN 61326-1:2013, group 1, class B
Electromagnetic
– IEC 61326-1:2012, group 1, class B
compatibility
– EN 61000-3-2:2019
– IEC 61000-3-2:2018
– EN 61000-3-3:2013+A1:2019
– IEC 61000-3-3:2013+AMD1:2017
– USA (FCC): 47 CFR, Part 15, Subpart B – Unintentional
radiator class B
– Canada (IC): ICES-003 – Unintentional radiator class B

Publicidad