All manuals and user guides at all-guides.com
Accuracy
See the Specifications section for operating conditions.
Accuracy is specified as follows: ± (a percentage of the reading + the value of the rightmost
decimal place) at 18 °C to 28 °C (64 °F to 82 °F), 0% to 80% relative humidity.
Accuracy Table
Value
Range
True Power
600.0 kW
Apparent Power
600.0 kVA
0.3 to 1.0
(
= 72.5° to 0°)
Power Factor
0.0 to 0.3
(
= 90° to 72.5°)
DCA
999.9 A
ACA††
999.9 A
Peak Current*
999.9 A
DCV**
600.0 V
ACV** ††
600.0 V
Peak Voltage***
600.0 V
999.9 Ω
Continuity
Resistance
9.999 kΩ
Frequency†
999.9 Hz
* Peak detect acquisition time ≤ 100 ms
** Input impedance is 1 MΩ.
*** Peak detect acquisition time ≤ 200 ms
† Sensitivity is 10 V or 10 A.
†† AC-coupled, true RMS measurement,
additional error for crest factors > 1:
Crest Factor
Error (% of reading)
1 to 3
0.5%
3 to 5
3%
5 to 7
6%
CMPL-100
Accuracy
Frequency Range
± (2% + 0.5 kW)
40 to 400 Hz
± (2% + 0.5 kVA)
40 to 400 Hz
± 3°
10 to 60 Hz
Not
Specified
± (2% + 0.5 A)
N/A
± (2% + 1.0 A)
40 to 400 Hz
± (2.5% + 1.0 A)
40 to 400 Hz
± (0.75% + 0.2 V)
N/A
± (1.2% + 1.0 V)
40 to 400 Hz
± (2.5% + 1.0 V)
40 to 400 Hz
± (1% + 0.5 Ω)
N/A
± (1% + 0.005 kΩ)
N/A
± (0.5% + 0.5 Hz)
0 to 1 kHz
19