Infratec™ Sofia
Message
BG20 test outside limits
Noise test failed
Noise test outside limits
Corrupt prediction data
Prediction failed
Could not start sequence
Could not access data in system
Could not use product. Corrupt file.
Too many parameters in total
No usable products
No Check Cell product found
HW module hang up
Temperature outside limits
Test sequence failed
Test sequence stopped by user
Wrong chassis ID
No valid Check Cell definition exists
Check Cell results outside limits
Check Cell should be analyzed
No update file present
Setting could not be saved
Too many UDFs
SHW not empty
Too many errors generated
User Manual 6002 5350 / Rev. 6
Action
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Reload file to SD-Card from FossManager
Updater and retry analysis.
Retry analysis. If problem persists, please con-
tact your local service representative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Corrupt file. Reload file to SD-Card from Foss-
Manager Updater.
Maximum number of parameters is 25. Correct
number of products and/or constituents in Foss-
Manager Updater.
Reload products to SD-Card from FossManager
Updater.
Reload Check Cell product to SD-Card from Fos-
sManager Updater.
Restart instrument and retry.
Wait for instrument to reach operating temper-
ature.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument.
Please contact your local service representative.
Run Check Cell definition sequence.
Restart instrument and run Check Cell analysis.
Run Check Cell analysis.
Reload file to SD-Card from FossManager
Updater.
Check write protection on SD-Card.
Reload from FossManager Updater and retry.
Run flush manually.
Please contact your local service representative.
6:3