5.3.4
Battery Type Selection
HR6 = rechargeable NiMH batteries, LR6 = non-rechargeable alkaline batteries
6.
Operation
Press
and hold for one second to turn the unit ON.
6.1
Clinical Use
Apply a liberal amount of water based ultrasound gel on the site to be examined. Place the probe at 45° to
the skin surface over the vessel to be examined. Adjust the position of the probe to obtain the loudest audio
signal. High pitched pulsatile sounds are emitted from arteries while veins emit a non-pulsatile sound similar to
a rushing wind. For best results, keep the probe as still as possible once the optimum position has been found.
Adjust the audio volume as required. Additional icons appear in the Setup mode.
6.1.1
Waveform Options
1. Freeze Live trace screen (refer to Frozen Trace section 6.3)
2. Switch between Venous and Arterial Modes
3. Switch between Forward and Reverse Flow
4. Select Timebase
Setup Screen Icons
Freeze Trace
Trace Options
Set Timebase