W&H Osstell Beacon Manual Del Usuario página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 84
The measured ISQ values will be displayed in the upper display for a
couple of seconds and then switch to indicate ready for measurement
in mesial-distal direction. See fig 5.
Note! Do not bring the instrument back in the mouth until the display
has switch to the next direction.
5. Repeat step 4 to measure in the mesial-distal direction, see fig. 6,
and then the sequence starts over and the Osstell Beacon is ready for
measurement in the BL (Buccal- Lingual) direction again. See fig 7.
6. When measurements in both directions are performed, remove the
SmartPeg by using the SmartPeg Mount.
7. T he instrument is automatically turned off after 60 seconds of no
motion.
1
3
5
7
25087-01 GB
4
BL
6
MD
8a
approx. 45º
3-5 mm
8b
8c
2
9

Publicidad

Tabla de contenido
loading

Tabla de contenido