EN-6
SJX (Internal Calibration Mode):
Press and hold Menu until [mMeNU] (Menu) is displayed. Release the
button, the display will show [C.A.L]. Press Yes to accept. [In.CAL] will
then be shown. Press Yes to begin the Internal calibration. The zero
reading is stored automatically. Pull the InCal handle while the
display show [PULL]. Push the InCal handle while the display show
[PUSH]. Repeat the Pull and Push steps one more time. The display
will show [done] if the calibration was successful. The balance
returns to the previous application mode and is ready for use.
3.4 Weighing Mode
This mode is the factory default setting.
1. If needed, press and hold Mode until [wWeIGH] (Weigh) is displayed.
2. If required, place an empty container on the pan and press Tare.
3. Add sample to the pan or container. The display shows the weight of the sample.
3.5 Counting Mode
This mode counts large numbers of items based on the weight of a reference count.
1. If required, place an empty container on the pan and press Tare.
2. Press and hold Mode until [Count] (Count) is displayed. [CLr.PwW] (Clear Average Piece
Weight, APW) will then display.
If no APW exists, the balance will display [PUt.10], proceed to step 5.
3. Press No to use the stored APW. Proceed to step 7.
4. Press Yes to establish an APW.
5. The balance will then display the sample size, i.e. [PUt 10]. Press No or Back to toggle the
choices (5, 10, 20, 50 or 100).
6. Put the indicated number of pieces on the pan then press Yes to calculate the APW. The
display shows the piece count.
7. Add additional pieces until the desired count is reached.
8. To clear the stored APW press and hold Mode until [Count] is displayed. Press Yes when
[Clr.pwW] is displayed.
Note: Press Function to view the current APW.
3.6 Percent Mode
This mode measures the weight of a sample as a percentage of a reference weight.
1. If required, place an empty container on the pan and press Tare.
2. Press and hold Mode until [Percnt] is displayed. [Clr.ref] (clear reference) will then
display. If no reference weight exists, the balance will display [Put.ref], proceed to step 5.
3. Press No to use the stored reference weight and proceed to step 6.
4. Press Yes to establish a new reference. Balance will now display [Put.ref].
5. Add the desired reference material to the pan or container. Press Yes to store the
reference weight. The display shows 100%.
6. Replace the reference material with the sample material. The display shows the
percentage of the sample compared to reference weight.
7. To clear the stored reference press and hold Mode until [Percnt] is displayed. Press Yes
when [Clr.ref] is displayed.
Note: Press Function to view the current reference weight.
TM
SCOUT
SJX SERIES BALANCES