Safety instructions
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return line or in the active circuit; this can lead
to the patient suffering burns or the outbreak of a fire.
If damaged, do not use these products!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during reconditioning and prevent fluids
from penetrating the connector during use. The older the mate-
rial, the greater the risk of unwanted damp in the connectors.
Erbe Elektromedizin expressly warns against modifying the ca-
ble, e.g. independently replacing defective insulation. Any mod-
ification leads to the termination of liability on behalf of Erbe
Elektromedizin.
5
How to use
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
WARNING! Do not kink the cable.
WARNING! Protect the product from all physical damage! Do not
throw! Do not apply force of any kind!
WARNING!