Processing Limitation; Dismantling; Required Aids; Precleaning - Erbe 20196-115 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58
OBJ_DOKU-184685-011.fm Seite 23 Freitag, 28. August 2020 8:49 08
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
6.2

Processing limitation

Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
6.3

Dismantling

1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
6.4

Required aids

Processing step
Items Needed

Precleaning

Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)
6.5
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

20196-11820196-11920196-124

Tabla de contenido