6.2
Processing Limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
6.3
Disassembly
1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
6.4
Required Aids
Processing step
Precleaning
Manual cleaning/
disinfection
6.5
Precleaning
Use tap water for precleaning, a non-fixing disinfectant if neces-
sary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
Items Needed
Soft plastic brush/soft cloth
Soft plastic brush/soft disposable
cloth (sterile, lint-free)
Compressed air gun (2 bar maxi-
mum)