CAUTION
Distance to communication devices. If the distance to HF communication devices (e.g.
mobile phones, Bluetooth devices, WiFi devices) is too small, interference with internal data
communications can result in unexpected
prosthesis behaviour. Therefore, keeping the following minimum distances to these HF commu-
nication devices is recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,...): 0.22 m
• B luetooth devices (third-party products not approved by Ottobock): 0.22 m
INFORMATION
This product may not be disposed of with regular domestic waste. Disposal that is not
in accordance with the regulations of your country may have a detrimental impact on
health and the environment. Please observe the information provided by the responsible
authorities in your country regarding return and collection processes.
4 Adjusting the MyoBock electrodes
INFORMATION
• Allow the patient to rest during the adjustment of the electrodes, otherwise muscle fatigue will
lead to inconsistent results and the therapist will, as a result, tend to adjust the electrodes
with too high a sensitivity.
• The electrodes should be placed on intact skin only and with as much electrode-skin contact
as possible.
• In the case of strong interference from electric devices, the position of the electrodes should
be verified and adjusted if necessary. If interference cannot be eliminated, please contact
Otto Bock Myo-Service.
• Before connecting, apply sufficient Silicone Grease 633F11 to the electrode pin contacts
and to the connecting cable contacts. The grease will protect the connection from humidity.
5 Scope of Delivery
1 pc. 7in1 Controller 9E420
1 pc. Contact Plate with cable 9E414
1 pc. Lamination Dummy 9E421
6 Accessories
1 pc. Lamination Ring 10S16
1 pc. Installation Tool 711M50-1
1 pc. Adapter Cable 757VAK2 for VASI Hand
1 pc. MyoEnergy Integral 757B25=1, 757B35=*
9E420=L, 9E420=R
Ottobock | 21