Note: If the instrument uses G1/8 connections, and the pressure is ≥ 1500 psi (100 bar),
connection method (b) is mandatory.
a) 1/8 NPT
connections
c) G1/8 connections:
Pressure < 1500 psi (100 bar)
1
Process connector. Maximum torque:
1/8 NPT: 26 lbf.ft (35 Nm)
G1/8: 18.4 lbf.ft (25 Nm)
3
(1/8 NPT only) Thread with an
applicable sealant
5
Reference port connector: M5 or 10-32
UNF (Table 10)
Maximum torque: 1.5 lbf.ft (2 Nm)
2.2 Communications Port Connections
Use the communications port (Figure A1 - item [9]) to attach an IDOS Universal Measurement
Module (UMM).
When you attach the cable from a UMM (Figure 8), the instrument automatically changes the
menus to give you all the applicable options (Table 2 & Table 3).
2.3 Measure Pressure
To measure the pressure or the pressure difference (if there are two IDOS pressure ranges).
1.
Connect the instrument (Figure 2) and, if necessary, adjust the Set Up (Table 1).
2.
Select an applicable pressure task from Select Task (Table 2/Table 3) and, if necessary,
adjust the Settings (Table 4).
3.
If necessary, do a zero correction (Table 4).
2
1
3
1
2
4
Figure 1: Connection Methods
b) G1/8 connections:
Pressure ≥ 1500 psi (100 bar)
1
4
d) M5 or 10-32 UNF
connections: Refer to
5
2
Applicable instrument pressure connector
(Table 9)
4
(G1/8 only) Applicable bonded seal
English–DPI 800/802 Instruction Manual | 11