When you select a channel, the display shows the applicable instructions to complete the
calibration.
When the calibration is complete, select Calibration Due and set the new calibration date for the
instrument.

4.2 Procedures (Pressure P1/P2)

1.
Connect the instrument to the pressure standard (Figure 2).
2.
Let the equipment get to a stable temperature (minimum: 30 minutes since the last power
on).
3.
Use the calibration menu (Table 6) to do a two-point calibration (Zero and +FS) or a three-
point calibration (-FS, Zero and +FS). Refer to Table 7. The display shows the applicable
instructions to complete the calibration.
Ranges: g/d
≤ 10.0 psi (700 mbar)
> 10.0 psi (700 mbar)
a. For a three-point calibration, do not apply more than -90% of the specified FS for the unit.
Ranges: a
5.00 psi (350 mbar)
Copyright 2003 Baker Hughes Company.
20 | DPI 800/802 Instruction Manual–English
Table 6: Calibration Options
Description
To calibrate the specified IDOS pressure input (P1 or P2
UMM only. To calibrate the specified IDOS
user manual for the IDOS UMM.
DPI 802/802P only. To calibrate the mA input.
Calibration Due: To set the date of the next calibration for the
instrument.
After the specified calibration date, there is a warning
message. There is a selection box to stop the warning.
To change the calibration PIN (Personal Identification
Number).
Table 7: Calibration Pressures
Nominal Applied Pressure
psi (mbar)
-FS
a
-FS
-13.1 (-900)
Nominal Applied Pressure
psi (mbar)
Zero
< 0.02 (1.0)
UMM
. Refer to the
Zero
+FS
0
+FS
0
+FS
+FS
+FS
).