10 Totalization
Tolerance control is possible in weighing or target mode.
When using in the counting system, no matter, if the weighed goods are on the
reference or on the bulk material balance.
Preparing:
When using as counting system select using
totalized. The appeared [] displays the current balance.
When totalizing in counting mode set the average piece weight.
(see chap. 8.1 or 8.2)
If necessary, tare the empty balance container.
10.1 Manual totalizing
With this function the individual weighing values are added into the summation
memory by pressing
• Menu settings:
„F1 off" „ACC" „ON" (unavailable in model CFS 50K-3)
„F2 Prt" „P mode" „Print" „Au OFF" (cf. chapter 12.2)
• When using as counting system there can be totalized on the reference as
well as on the bulk material balance.
Before the totalizing process select the active balance, see chap. 7.3.
Add up:
Place weighing goods A.
Wait until stability display is displayed, then press
3). The weight value or number will be stored and be output if connected to an
optional printer.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place goods to be weighed B.
Wait until stability display is displayed, then press
3). The weight value or the number of items are added in the summation memory
and printed out. Total weight, number of weighing procedures as well as total
parts counting appear 2 sec.
Add more weighed goods as described before. Please note that the balance must
be unloaded between the individual weighing procedures.
CFS/CCS-BA-e-1523
and edited, when an optional printer is connected.
the balance, on which will be
or
(models CFS 50K-
or
(models CFS 50K-
41