Use the numeric keys to input password „9999":
Confirm input by
Select bulk material or reference balance via
The appeared [] displays the current balance.
When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out
on both balances.
Acknowledge with
Using
select the weighing unit [kg or lb] which will
be used for linearization. The appeared [] displays the
current weighing unit. Acknowledge with
Ensure that there are no objects on the weighing pan.
Wait for stability display (indicator [] atop
appears), then press
Carefully place first adjustment weight into the center of
the weighing plate on display „LoAd 1".
Wait for stability display, then press
Carefully place second adjustment weight into the
center of the weighing plate on display „LoAd 2".
Wait for stability display, then press
When „LoAd 3" is displayed, place the third adjustment
weight carefully in the centre of the weighing plate.
Wait for stability display, then press
After the adjustment the balance will carry out a self-
test. The balance returns automatically into weighing
mode. In case of an adjustment error or incorrect
adjusting weight the display will show an error message
(
CFS/CCS-BA-e-1523
.
.
.
), repeat adjustment process.
„Pin" „----„
.
„tECH" „LoCAL"
„tECH" „rEmotE"
„tECH" „Unit"
.
.
.
.
69