Alarm Conditions

All alarms are low priority alarms.
Alarm
General malfunction
yellow light
and intermittent
audible alarm
Oxygen monitor yel-
low light
and
intermittent audible
alarm
Power failure yellow
light
and
intermittent audible
alarm
Indicates
high product tank pressure
OR
low product tank pressure
OR
high device temperature
low oxygen concentration
power failure
Action
Ensure flowmeter is open to minimum flow rate or higher.
Ensure cannula is not kinked or obstructed. Remove any
devices connected downstream of the outlet of the device.
Ensure device has at least 12" of clearance on all sides
and intakes are not obstructed. Ensure external gross
particle intake filter is clean and not clogged. Ensure unit
is within operating temperature range. If issue persists,
contact equipment provider for service.
Contact equipment provider for service.
Ensure device is plugged into a known, working outlet. En-
sure breaker switch is pushed in. If issue persists contact
equipment provider for service.
PN MN238-C4 E | User Manual US - 17