EMC Testing

Medical equipment needs special precautions regarding EMC and needs to be installed and put into ser-
vice according to the EMC information provided in this section.
Guidance and Manufacturer's Declaration —Electromagnetic Emissions
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of
the VisionAire should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF
communications equipment and the VisionAire Units
The VisionAire is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the VisionAire can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the VisionAire as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in me-
tres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
20 - US PN MN238-C4 E | User Manual
Compliance
Group 1
Class B
Complies
Complies
Separation distance according to frequency of transmitter
from 150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
Electromagnetic Environment - Guidance
The VisionAire uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The VisionAire is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
m
from 80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
from 800 MHz to 2,5 GHz
d
1
2 .
P
0.23
0.73
2.3
7.3
23