CARE AND MAINTENANCE

Surface Care
As needed, use a damp cloth or sponge, with mild detergent if necessary. Do NOT use abrasives
or solvents, as these may degrade the case. Do NOT submerge the LOCalizer Reader in liquid.
Batteries
Periodic battery replacement is necessary for the unit to continue to operate correctly. To replace
batteries, open the battery compartment by holding the textured edges and pulling firmly off the
device. Install commercially available AA batteries following the battery polarity symbols on the inside
of the case. Approximate battery life expectancy under normal use conditions is 8 hours. These times
may be slightly reduced at high display intensity and audio volume. Use only good quality AA Alkaline
cells (IEC-LR6).
REPLACEABLE / REPAIRABLE COMPONENTS
There are no repairable components. With the exception of battery replacement, (Refer to Care
and Maintenance) no modification or repair of this equipment by the user is allowed.
If the LOCalizer Reader stops working or is damaged it should be replaced.

TECHNICAL INFORMATION

Environmental, Storage, and Transportation Conditions
LOCalizer Reader
Ambient Temperature
Relative Humidity
Atmospheric Pressure
*For best service life of AA Alkaline Cells used in LOCalizer Reader, storage temperature should be
5°- 30°C. To avoid potential damage from cells leaking, remove battery if the LOCalizer Reader will
not be used within 2 months.
Store the Tag Applicator and Surgical Probe at ambient temperature 15° to 30°C (59° to 86°F).
Interference
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause
undesired operation.
The user is responsible to maintain FCC compliance. Any changes or modifications not expressly
approved by Health Beacons could void the user's authority to operate the equipment.
Use the LOCalizer Reader only with a LOCalizer Surgical Probe and Tag.
Operational Conditions
5˚ - 30˚C (41˚ - 86˚F)*
30% - 75%, non-condensing
70 kPa – 102 kPag
Temporary Storage and
Transport Conditions*
-20° - 50°C (-4°-122°F)
15% - 93%, non-condensing
50kPa - 106kPa
User Manual 11