leak tightness of the pressure
system and possible deviations of
the pressure reading.
Remove at least one battery in
order to switch to calibration mode.
Hold down on the START/STOP but-
ton and then insert the battery.
Release the button and, after a few
moments, two zeros will appear
one above each other on the dis-
play.
Instructions on the measurement
function check will be supplied on
request to competent authorities or
authorised maintenance providers
by HARTMANN.