9. Explanation of error displays

Error which has
occurred
Device will not turn
on.
Cuff will not inflate.
All manuals and user guides at all-guides.com
Possible causes
No batteries, they have
been inserted incorrect-
ly or are dead.
Cuff defective.
Moving or talking
during the measuring
procedure.
Cuff is too loosely
applied.
Cuff loses air/has a
leak.
Measurement
inaccurate.
The pressure in cuff
exceeds 300 mmHg.
Therefore, an automatic
fall in pressure occurs.
94
Remedy
Check batteries and
insert two identical, new
batteries if necessary.
Return the device to the
designated service
center for inspection.
Do not talk or move
during the measuring
procedure.
Apply the cuff so that it
fits snugly around the
wrist.
Return the device to the
designated service
address.
Please rest for at least a
minute and take the
measurement again.
Please rest for at least a
minute and take the
measurement again.