Descargar Imprimir esta página

TECNOMOTOR TM 514 Manual De Instrucciones página 239

Ocultar thumbs Ver también para TM 514:

Publicidad

Idiomas disponibles

Idiomas disponibles

Produzido por Tecnomotor Eletrônica do Brasil S/A - REPRODUÇÃO PROIBIDA - Eventuais erros ou defeitos favor comunicar tecnomotor@tecnomotor.com.br
Error
179
Code
Probable Cause
- Bad contact on sensor(s) connections.
Error
180
Code
Probable Cause
- Bad contact on sensor(s) connections.
Error
192
Code
Probable Cause
- Injector in trouble.
- Occurrence of error 195/196.
Error
193
Code
Probable Cause
- Jet breaker adapter disconnected.
- Leak in drain valves.
- 3 way valves disconnected.
- 3 way valves with leaks.
Error
194
Code
Probable Cause
- Jet breaker adapter disconnected.
- Leak in drain valves.
- 3 way valves disconnected.
- 3 way valves with leaks.
Error
195
Code
Probable Cause
- Accumulation of impurities in the hydraulic
system.
- Bad contact at the drain valve connections.
- Damaged power board.
- Blown valve fuse.
Error
196
Code
Probable Cause
- Accumulation of impurities in the hydraulic
system.
- Bad contact at the drain valve connections.
- Damaged power board.
- Blown valve fuse.
TM 514 – COMMON RAIL TEST
CH1 optical module disconnected
- Contact the manufacturer.
CH2 optical module disconnected
- Contact the manufacturer.
Direct injection tightness test failed
- Contact the manufacturer.
CH1 measurement timeout
- Check jet breaker connections.
- Check actuation of 3 way valves.
- Check for leaks in 3-way valves.
- Check valve fuse.
CH2 measurement timeout
- Check jet breaker connections.
- Check actuation of 3 way valves.
- Check for leaks in 3-way valves.
- Check valve fuse.
Impossible to empty the meter CH1
- Check valve fuse.
- Apply 'auto clean' functionality to production
test actuators.
Impossible to empty the meter CH2
- Check valve fuse.
- Apply 'auto clean' functionality to production
test actuators.
237
Solution
Solution
Solution
Solution
Solution
Solution
Solution

Publicidad

loading