If the InBody570 is not plugged into a grounded outlet, it may cause damage through electric surges or
•
product malfunction. This may affect the test results.
The test results may be inaccurate if the InBody570 is under electrical interference. Do not install the InBody570
•
near products that generate electrical interference such as fluorescent lights, large AC motor equipment
(treadmill, vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal therapy
equipments, or heating appliances. Do not share the power source of the InBody570 with other electrical
devices. This may affect the test results.
When connecting the InBody570 with other test equipment, turn on the other equipment first. When
•
turning off other equipment, turn off the InBody570 first. This is necessary to minimize electrical surges on the
InBody570.
Always use the specified adapter provided by InBody as it is a part of the InBody570. Using other adapters
•
may result in malfunction of the InBody570.
Operation of the InBody570 2,000m above sea level may affect the weight measurement.
•